|
Cleaning, Maintenance and
Resin Removal
There
are two times cleaning must to be performed to Porcerax
II:
The
first time is when the tool shop has prepared the insert
or cavity. The material is 25% air by volume and the cutting fluids will displace
the air, thus filling and clogging the pores. After the permeability has been restored, by
either stoning or EDMing the venting surface, the steel should be placed in an oven at about
350°F. for approximately 2-3 hours. This will thin and remove most of the fluids in
the pore structure. Once cooled to room temperature,
place the insert in an ultrasonic cleaning unit (built
for flammable solvents) filled with acetone. These units have spark-proof switches,
water chillers and are temperature monitored. This is the only recommended type
cleaner for cleaning Porcerax II.
The
steel should be thoroughly blown dry through a drilled,
tapped, and EDM-relieved hole in the side of the insert or cavity. This clean-out
or back-flush hole should be designed to accept a shop air fitting. This will be very useful
over the life of the insert in back-flushing fluid and future residue from the insert, and for
checking restored permeability with filtered shop air. (See photo below.) If the cleaning operation
is not completed, due to the trapped fluids, inside permeability will not exist.
Note: In addition to
the other cleaning methods, it may be necessary to
back-flush the Porcerax II inserts with acetone, using
air pressure, in order to thoroughly clean the inserts.
Care should be taken when using solvents under pressure.
Note:
These cleaning steps may need to be repeated
Note:
Do not use any heating device that may be supplied by
the ultrasonic cleaner.

The
second time cleaning is needed is after the insert or
cavity has been in operation and film, a thin layer of mold release agent, resin residue,
shop oil or any other contaminate has partially or totally blocked the air flow. There are
two primary ways to clean this surface:
- Spray
or brush a liberal amount of acetone, mold cleaner
or solvent onto the contaminated surface of the steel, wait a few minutes, then
back-flush the insert with filtered compressed air. This can be done through the drilled and tapped
hole previously discussed. Or, if material is inserted, by tapping the exhaust line
that exits the core block and attaching an air fitting there.
- The
second and preferred way to clean the porous
material is to remove the insert or take the small cavity and place the insert in an
ultrasonic cleaning unit (built for flammable solvents) filled with acetone. These units
have spark-proof switches, water chillers, and are temperature mounted. This
is the only recommended type cleaner for cleaning Porcerax II.
Note: In addition to
the other cleaning methods, it may be necessary to
back-flush the Porcerax II inserts with acetone, using
air pressure, in order to thoroughly clean the inserts.
Care should be taken when using solvents under pressure.
Note:
Either of these cleaning steps may need to be repeated
Note:
When cleaning Porcerax II, if Dichtol sealed waterlines
are present, the need to reseal these waterlines will be required.
Note: With some
engineered resins it may be necessary to clean Porcerax
II inserts with other chemicals such as a 6 to 8% Sodium
Hydroxide solution. This can be used in an ultrasonic
cleaner with temperatures up to 160 degrees F, for a
period not to exceed 40 minutes. Immediate flushing with
water as a neutralizer should take place in order to
prevent corrosion within the Porcerax II insert. A
45-minute ultrasonic cleaning in water should follow
this. Then remove the insert and dry with filtered
compressed air. Consult your chemical supplier for the
necessary safety instructions!
For further
information and questions regarding Porcerax II cleaning
please call IMS: 1-800-625-6653
|